page contents

ECL101 & ECL110

Eddy-Current Displacement Sensor
ECL101 – Single Channel DIN Rail Mount
ECL110 – Multi-Channel (No Enclosure)

Linear output at up to 80 kHz bandwidth. Can be calibrated in the field.

Download Documentation:

 

ECL101 User Manual

ECL110 User Manual

clear label sensor catalogEddy-Current Sensor Catalog

label sensor CE documentCE Declaration of Conformity

Description

Precision Measurements

Easy to Use

  • Front Panel Adjustable Zero (Offset)
  • Field Calibration Adjustments
  • Range Indicating LEDs
  • 12-24 VDC Power
  • Probes can be ordered as vacuum compatible.

Life-Cycle Status: Active

Applications

Thread Sensing in Nonferrous Materials

ecl110 eddy current sensor ECL110

ECL110 Multi-Channel Version

The ECL110 is a multi-channel, OEM version of the ECL101. The ECL110 has no enclosure, is powered by a single power connection, and can have 1-8 channels internally synchronized with no need for external sync cables.

Specifications

Typical specifications for standard targets and probes. Actual values depend on probe, target, and calibration. Specifications assume a target area at least three times the diameter of the probe.

Resolution: See Range Table for resolution specifications
Linearity 0.50% of Range
Bandwidth (Factory Set) 1 kHz, 10 kHz, 80 kHz
Temperature Drift of Probe at mid-range
(15°C-65°C)
0.04% of Range/°C
Output 0-10 VDC
Input Power 12-24 VDC

130 mA at 15 VDC

Operating Temperature 4°C-50°C
Probe Operating Temperatures

Standard Probes come with PUR (Polyurethane) jacketed cables. High Temperature Probes have FEP jackets.

Other temperature ranges may be possible. Contact us for more information.

Standard Probes: IP67

High Temperature Probes: IP63

U3 Probe Standard Probe (PUR Cable):
-25°C to +125°C
(Available with optional FEP cable)High Temperature Version Not Available
U5-18 Probes Standard Probes:
-25°C to +125°CHigh Temperature Probes:
-25°C to +200°C
U25-50 Probes Standard Probes:
-25°C to +125°CHigh Temperature Probes:
-25°C to +175°C

 

Probes, Ranges, and Resolutions (Probe Dependent)

Listed specifications assume a 3 meter probe cable; flat measurement area diameter at least 3 times larger than the probe diameter; and no customizations.

Probes can be ordered as vacuum compatible; custom probesare available.

cal

Listed specifications are for standard target materials: Nonferrous Al 6160, Ferrous 4140 Steel.

Range

mm
inch

Near Gap

mm
inch

Probe

probe flyout

Target
Type

Resolution (RMS*)

µm
µinch

1 kHz 10 kHz 80 kHz
0.50
0.020
0.05
0.002
U3

eddy-current sensor probe

Nonferrous 0.03
1.2
0.06
2.4
0.20
8.0
Ferrous 0.04
1.6
0.08
3.2
0.30
12
1.25
0.050
0.25
0.010
U5eddy-current sensor probe Nonferrous 0.06
2.4
0.10
4.0
0.25
10
Ferrous 0.09
3.6
0.15
6.0
0.40
16
2.00
0.080
0.35
0.015
U8eddy-current sensor probe Nonferrous 0.10
4.0
0.16
6.4
0.40
16
Ferrous 0.13
5.2
0.21
8.4
0.50
20
3.50
0.140
0.60
0.025
U12eddy-current sensor probe Nonferrous 0.20
8.0
0.28
11
0.70
28
Ferrous 0.26
10
0.35
14
1.2
48
5.00
0.200
0.75
0.030
U18eddy-current sensor probe Nonferrous 0.24
10
0.48
19
3.2
130
Ferrous 0.32
13
0.64
26
4.5
180
8.00
0.320
1.25
0.050
U25

eddy-current sensor probe

Nonferrous 0.35
14
0.70
28
5.3
210
Ferrous
12.5
0.500
1.50
0.060
U38eddy-current sensor probe Nonferrous 0.55
22
1.1
44
8.3
330
Ferrous
15.0
0.600
2.00
0.080
U50eddy-current sensor probe Nonferrous 0.66
26
1.3
52
10
440
Ferrous
* Peak-to-Peak values are approximately 8-10 times RMS values. ECL101 is ordered and shipped with the desired bandwidth. In high EMI environments (10 V/m), output noise may rise to 0.050VRMS (0.5% resolution)

 

Mechanical

ECL101

sd

ECL110

ECL110 Outline Drawing